Portable syngas analyzer for long-term measurements of special gases (syngas)
Within most compact space one electrochemical O2 sensor (or one paramagnetic O2 sensor), one thermal conductivity detector (H2) as well as an infrared bench (CO+CO2+CH4) work jointly to generate precise measuring results.
All important interfaces are available, such as Ethernet (LAN), WLAN, Bluetooth, USB, RS485, 8 channel analog outputs
Remote wireless control using a smartphone and MRU4u app
Benefits:
Make your combustion process more cost-effective.
Get the data to show your organization cares about the environment.
Applications:
Glass Industry, Iron / Steel Industry, Process / Syngas Analysis, Universities / Laboratory
– O2 sensor (either electrochemical longlife sensor or paramagnetic sensor)
– Integrated, electrical gas cooler (Peltier) and automatic condensate draining pump
– Strong sample gas pump and external accessible particulate Teflon filter
– Automatic zeroing by means of 3-way solenoid valve, user programmable
– Internal sample flow monitoring with display and alarm
– Linux OS and large baklit 840 x 480 colour display, with touch and swipe technigue
– With universal analog input (AUX socket) for external transmitter connection
– Data memory for approx. 10.000 measurements, optional USB Stick for additional data storage
– Automatic interval measurement program
– Data-visualization and evaluation software MRU4win
– Universal power supply: mains 90-265 Vac /105 W
– Internal Li-Ion battery for 1 hours mains free operation
– Automatic selftest and control of soft- and hardware functions
– Modern, rugged aluminium enclosure with rubber molded impact protection
Gas measurement
Method
Measuring range min./max.
Resolution
Accuracy
O2
Oxygen (Long Life)
EC
0 … 25 %
0.01%
0.20%
O2
Oxygen
PM
0 … 25 %
0.01%
0.1%
CO
Carbon monoxide
NDIR
0 … 10.00 % / 100.00 %
0.01%
± 0.1 % or 2 % reading
CO2
Carbon dioxide
NDIR
0 … 10.00 % / 100.00 %
0.01%
± 0.3 % or 2 % reading
CH4
Methane
NDIR
0 … 10.00 % / 100.00 %
0.01%
± 0.2 % or 2 % reading
H2
Hydrogen
TCD
0 … 10.00 % / 100.00 %
0.01%
± 0.1 % or 2 % reading
H2S
Hydrogen sulfide
EC
0 … 2,000 ppm / 5,000 ppm
1 ppm
± 5 ppm or 5 % reading
Calculated components
Calorific value N2 balance
O … SO MJ/m3 or MJ/kg difference to 100%
Other measurements
Method
Measuring range
Resolution
Accuracy
Stack gas temperature (Tgas)
NiCrNi
0 … 2,012 °F
1 °F
± 4 °F or 2 % reading
Combustion air temperature (Tair)
NiCrNi
0 … 212 °F
1 °F
± 2 °F or 1 % reading
Differential pressure (P-Druck)
Piezoresistive
–48 … +48 inH2O
0.001 inH2O
± 0.008 inH2O or 1 % reading
Flow velocity measurement (v)
Pitot
3 … 100 m/s
0.1 m/s
± 1 m/s or 1 % reading
Standardized ext. signal (AUX connection)
Software
for K-thermocouple, 0 … 10 Vdc, 4 … 20 mA, RS 485
Combustion calculations (fuel type depend.)
Software
Losses, Excess Air, Air Ratio, dew point, CO₂
Emissions calculations
Software
mg/Nm3, reference to O₂
General technical data
Operating system
LINUX
Display, operation
7″ TFT (800 x 480 px) color display, backlit, with touch pad
Data storage type
dynamic, internally 10,000 data sets, external USB stick