SWG 300 – Continuous Emission Monitor

SWG 300 CEMS Analyzer OUTDOOR
SWG 300 CEMS Analyzer EX
SWG 300 CEMS Probes 002
SWG 300 CEMS Probes 003
SWG 300 CEMS Probes 004
SWG 300 CEMS Probes 005
Complete CEM system for emissions monitoring
  • dual heat exchanger M&C Peltier gas cooler with double automatic condensate draining pumps
  • Efficient sample gas filtration using PTFE sintered particulate filter
  • Condensate monitoring and alarm on display and gas sampling stop in case of alarm
  • Filtration of acid gases to protect the internal sample gas flow sensor

Description

Features:

  • Complete CEM system for emissions monitoring
  • The multi-component gas analyzer SWG 300-1 is based on extractive, cold-dry method
  • The SWG 300 can measure up to 8 gas components, using MRU optimized NDIR technology, with improved accuracy and without zero offset.
  • Oxygen analysis using, paramagnetic cell or “long-life” electro-chemical cell.

Available as:

  • SWG 300 INDOOR
  • SWG 300 OUTDOOR
  • SWG 300 EX-ZONE 2

Benefits:

  • High End Stationary Continuous Emission Monitoring system.
  • Improved accuracy and without zero offset.

Applications:

Coal power stations, Municipal and other waste incinerators, Cement outlet stacks, Glass melting industry, Solid fuels and heavy oil combustion sites, Bagasse and biomass steam boilers, and others.

 

Additional information

Auto Cal

2/3/4x auto.cal.

Enclosure Dimensions (H x W x D)

1012x600x575 / 1250x1000x800

Enclosure Material

Steel / Glass Fiber

Gas Cooler

Double Heat Exchanger M&C

Number of Heated Ports

4 (1 for "Ex-pz" model)

Number of Unheated Ports

n.a.

  • Use of optimized NDIR technology with improved accuracy and without zero offset
  • O₂ measurement with a long-life electrochemical or a paramagnetic sensor
  • True NOx and SO₂ measurements through use of heated gas sampling probe and gas sampling line
  • Heated and temperature regulated acid mist catch & drain system for raw flue gas measurements with high acid aerosols
  • Complete heavy duty sample conditioning system for extractive cold and dry system
  • Ready to log, display and transfer data from dust monitoring unit DM 401 and flow rate measuring unit DF 252.
Gas measurement (NDIR) Measuring range min./max. Resolution Repeatability * 8h-Drift * Linearity
NO Nitric oxide 0 … 200 / 4,000 ppm 0.1 ppm 2 ppm or 1 % reading 2 ppm or 1 % reading 1 % range
NO2 Nitric dioxide 0 … 150 / 1,000 ppm 0.1 ppm 1 ppm or 1 % reading 2 ppm or 1 % reading 1 % range
SO2 Sulfur dioxide 0 … 150 / 4,000 ppm 0.1 ppm 2 ppm or 1 % reading 2 ppm or 1 % reading 1 % range
CO2 Carbon dioxide 0 … 40% 0.01% 0.2 % or 1 % reading 0.2 % or 1 % reading 1 % range
CO Carbon monoxide 0 … 175 / 10,000 ppm 0.1 ppm 2 ppm or 1 % reading 2 ppm or 1 % reading 1 % range
N2O Nitrous dioxide 0 … 100 / 500 ppm 0.1 ppm 2 ppm or 1 % reading 2 ppm or 1 % reading 1 % range
CH4 Methane 0 … 500 / 10,000 ppm 0.1 ppm 10 ppm or 1 % reading 2 ppm or 1 % reading 1 % range
C3H8 Propane 0 … 200 / 5,000 ppm 0.1 ppm 2 ppm or 1 % reading 2 ppm or 1 % reading 1 % range
Gas measurement (EC/PM) Methode Measuring range min./max. Resolution Accuracy
O2 Oxygen (Long Life) EC 0 … 25 % 0.01% +/- 0.20% absolute
O2 Oxygen PM 0 … 25 % 0.01% +/- 0.1%
H2S Hydrogen sulfide EC 0 … 2,000/5,000 ppm 1 ppm +/- 5 ppm or 5 % of reading
H2 Hydrogen EC 0 … 1,000/2,000 ppm 1 ppm +/- 5 ppm or 5 % of reading
General technical data  
Zero offset negligible due to automatic zeroing
Span offset less than 0.2 % of the measuring range per month
Calculated components NOx: NO + NO₂, calculated ppm or mg/m3, user-selectable O₂ reference
Operation/interfaces
  • Backlit 3.5″ TFT color display
  • keyboard, password-protected operation
  • 8 analog outputs 4 … 20 mA, galvanically isolated, max. load: 500 R
  • 4 alarm relays, potential-free contacts: 24 Vdc, 5 A
  • Data storage and data logger on SD card
  • RS 485 digital interface (Modbus RTU)
  • DIN rail RS 485, to ProfiBus converter or to Ethernet converter
Gas conditioning
  • HD gas sampling probe, heated ceramic filter with backpurge,or gas sampling probe HD-GW, heated glass wool filter
  • 10 % H₃PO₄ dosage at low NOx and low SO₂
  • Heated acid separator with acid condensate pump
  • Heated gas sampling line, PTFE DN 4/6 mm, up to 50 m length
  • Thermoelectrical gas cooler, with dual heat exchanger and constant +4 °C dewpoint
  • Teflon-particle filter, internal Viton-tubing
  • Monitored and regulated gas sampling pump
  • Constant gas flow of 50 l/h
  • Gas inlet pressure: –200 … +50 mbar (hPa)
  • Sample gas outlet: atmospheric pressure
Operating conditions +5 … +45 °C or –10 … +45 °C with cabinet heating
Power supply Universal: 90 … 240 Vac, 47 … 63 Hz, 200 W (700 W with heating)
Protection class
   SWG 300 INDOOR IP54 steel cabinet
   SWG 300 OUTDOOR IP65 glass fiber re-enforced polyester cabinet
   SWG 300 EX ZONE 2 IP65 glass fiber cabinet
Dimensions (W x H x D)
   SWG 300 INDOOR 23.61” x 39.83” x 22.63” (600 x 1.012 x 575mm)
   SWG 300 OUTDOOR 39.35” x 49.19” x 31.48” (1.000 x 1.250 x 800mm)
   SWG 300 EX ZONE 2 49.19” x 39.35” 31.48” (1250x1000x800 mm)

 